Z-1633-2026 Class II Ongoing
FDA device recall Z-1633-2026 was initiated by Straumann USA LLC on Mar 17, 2026 and is designated Class II. Reason for recall: A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as… The recall status is ongoing. Affected quantity: 7 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 1, 2026
- Initiation Date
- Mar 17, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product description
Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
Reason for recall
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.
Code information
Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1633-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026