Z-1633-2026 Class II Ongoing

Recalled by Straumann USA LLC — Andover, MA

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
March 17, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Reason for Recall

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

Code Information

Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;