Z-1669-2014 Class II Terminated

Recalled by Synthes, Inc. — West Chester, PA

FDA device recall Z-1669-2014 was initiated by Synthes, Inc. on Apr 29, 2014 and is designated Class II. Reason for recall: Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use. The recall status is terminated (terminated May 19, 2016). Affected quantity: 6690.

Recall details

Product Type
Devices
Report Date
Jun 4, 2014
Initiation Date
Apr 29, 2014
Termination Date
May 19, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6690

Product description

Synthes ProDisc-C 2.0mm Milling Bit, Synthes 2.0mm Milling Bit HXC, Synthes 2.0mm Milling Bit with 2.35mm x 33.5mm Shaft, and Synthes 2.0mm Milling Bit with Flat Coupling. Spinal surgical instrument: The Synthes ProDisc-C 2.0mm Milling Bits are used as a part of the milling option for keel preparation of the vertebral bodies. The milling bits are coupled to a power tool and interface with the milling guide, which is connected to a trial implant for the keel preparation step.

Reason for recall

Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use.

Distribution pattern

US Nationwide Distribution

Code information

part number: 03.820.161s, with lot numbers: 5742507,5755970,5755971,5755972, 5755973,5755974,5828298,6336115, 6338099,6338100,6340029,6340030, 6344199,6344200,6363601,6385892, 6407046,6431260,6431261,6460402, 6503360,6526127,6571070,6571073, 6571074,6601985,6636946,6636947, 6636948,6653819,6653820,6670383, 6691341,6691342,6718933,6718934, 6718935,6740611,6794651,6794652, 6794653,6794654,6794655 part number: 03.820.1638 with lot numbers: 5742506,5755978,5755979,5755980, 5755981,5755982,5828299,5828300, 6320586,6336108,6336110,6336111, 6340033,6344201,6344202,6363602, 6385893,6431262,6431263,6460403, 6503361,6571075,6601986,6601987, 6636949,6636950,6636951,6653810, 6653818,6670382,6691343,6691344, 6718938,6740615,6740616,6745821, 6770474,6770475,6770476,6770477, 6770478,6794657,6794658 part number: 03.820.167S with lot numbers: 5755983,5755984,5755987,5755988, 5755989,5755985,5755986,5755990, 5828302,6338101,6338102,6338103, 6338104,6338106,6338107,6338108, 6338109,6344205,6471089,6670381, 6718939,6718940,6718941,6843907, 6843906,6856252,6856253,6867662, 6867661,6901712,6901713,6919726, 6919727,6919728,6901710,6901711, 6921717,6921718,6940357,6960192, 6960193,6983033,7006496,7055427 part number: 03.820.169S with lot numbers: 6066520,6113671,6113633,6159883, 6180175,6338068,6338098,6344206, 6363603,6407047,6431264,6460404, 6526128,6571078,6670384,6745820, 6770479,6770480

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1669-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Synthes, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1669-2014. Report a data issue: hello@regdatalab.com.