Z-1692-2021 Class II Terminated

Recalled by Draeger Medical, Inc. — Telford, PA

FDA device recall Z-1692-2021 was initiated by Draeger Medical, Inc. on Apr 16, 2021 and is designated Class II. Reason for recall: Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2… The recall status is terminated (terminated Feb 6, 2024). Affected quantity: 2 units.

Recall details

Product Type
Devices
Report Date
Jun 9, 2021
Initiation Date
Apr 16, 2021
Termination Date
Feb 6, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product description

Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

Reason for recall

Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning

Distribution pattern

US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.

Code information

All lots/serial numbers manufactured from 1/1/2019-3/31201

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1692-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Draeger Medical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Draeger Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Draeger Medical, Inc. →

Source: openFDA record for Z-1692-2021. Report a data issue: hello@regdatalab.com.