Z-1703-2020 Class II Terminated

Recalled by Sysmex America, Inc. — Lincolnshire, IL

FDA device recall Z-1703-2020 was initiated by Sysmex America, Inc. on Mar 9, 2020 and is designated Class II. Reason for recall: Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or… The recall status is terminated (terminated Mar 24, 2021). Affected quantity: 93 units.

Recall details

Product Type
Devices
Report Date
Apr 22, 2020
Initiation Date
Mar 9, 2020
Termination Date
Mar 24, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
93 units.

Product description

The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.

Reason for recall

Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.

Distribution pattern

USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..

Code information

Catalog No. AH699752.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1703-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1703-2020. Report a data issue: hello@regdatalab.com.