Z-1113-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 12, 2025
- Initiation Date
- January 6, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32 units
Product Description
Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.
Reason for Recall
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Distribution Pattern
US Nationwide.
Code Information
UDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967