Z-2770-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 21, 2016
- Initiation Date
- July 1, 2016
- Termination Date
- February 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3868 units
Product Description
Fluorocell PLT reagent, Catalog #CD994563 Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
Reason for Recall
There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) value and a false low PLT-F value. In most cases, the falsely decreased PLT-F results displayed a "PLT Abn Scattergram" message with asterisks beside the results indicating the data is unreliable. In rare instances, platelet flags may be absent. The PLT-I counts are unaffected. The PLT-F is a reflex test used to confirm PLT-I results due to flagging or low values. As such, the likely impact to patient results is low. Affected results display a discrepancy between the PLT-I and PLT-F, with lower than expected PLT-F results that do not match the patient's clinical picture.
Distribution Pattern
Worldwide Distribution - US (Nationwide, Washington DC) Canada; Cuba; Chile; and Dominican Republic.
Code Information
Lot number A5053, Expiration date 12/21/2016 Lot number A5054, Expiration date 12/21/2016 Lot number A5041, Expiration date 9/2/2016 Lot number A5042, Expiration date 9/2/2016 Lot number A5044, Expiration date 10/2/2016 Lot number A5046, Expiration date 11/9/2016 Lot number A5047, Expiration date 11/17/2016 Lot number A5048, Expiration date 11/17/2016 Lot number A5049, Expiration date 11/26/2016 Lot number A5050, Expiration date 11/26/2016 Lot number A5055, Expiration date 12/24/2016