Z-2770-2016 Class II Terminated

Recalled by Sysmex America Inc — Lincolnshire, IL

FDA device recall Z-2770-2016 was initiated by Sysmex America Inc on Jul 1, 2016 and is designated Class II. Reason for recall: There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue… The recall status is terminated (terminated Feb 10, 2017). Affected quantity: 3868 units.

Recall details

Product Type
Devices
Report Date
Sep 21, 2016
Initiation Date
Jul 1, 2016
Termination Date
Feb 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3868 units

Product description

Fluorocell PLT reagent, Catalog #CD994563 Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers

Reason for recall

There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) value and a false low PLT-F value. In most cases, the falsely decreased PLT-F results displayed a "PLT Abn Scattergram" message with asterisks beside the results indicating the data is unreliable. In rare instances, platelet flags may be absent. The PLT-I counts are unaffected. The PLT-F is a reflex test used to confirm PLT-I results due to flagging or low values. As such, the likely impact to patient results is low. Affected results display a discrepancy between the PLT-I and PLT-F, with lower than expected PLT-F results that do not match the patient's clinical picture.

Distribution pattern

Worldwide Distribution - US (Nationwide, Washington DC) Canada; Cuba; Chile; and Dominican Republic.

Code information

Lot number A5053, Expiration date 12/21/2016 Lot number A5054, Expiration date 12/21/2016 Lot number A5041, Expiration date 9/2/2016 Lot number A5042, Expiration date 9/2/2016 Lot number A5044, Expiration date 10/2/2016 Lot number A5046, Expiration date 11/9/2016 Lot number A5047, Expiration date 11/17/2016 Lot number A5048, Expiration date 11/17/2016 Lot number A5049, Expiration date 11/26/2016 Lot number A5050, Expiration date 11/26/2016 Lot number A5055, Expiration date 12/24/2016

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2770-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Sysmex America Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2770-2016. Report a data issue: hello@regdatalab.com.