Z-2953-2018 Class II Terminated

Recalled by Sysmex America, Inc. — Lincolnshire, IL

FDA device recall Z-2953-2018 was initiated by Sysmex America, Inc. on Jul 27, 2018 and is designated Class II. Reason for recall: The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT. The recall status is terminated (terminated Jul 8, 2019). Affected quantity: 5,168 units.

Recall details

Product Type
Devices
Report Date
Sep 5, 2018
Initiation Date
Jul 27, 2018
Termination Date
Jul 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,168 units

Product description

Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL. Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers.

Reason for recall

The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.

Distribution pattern

Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Argentina, Brazil, Canada, Chile, Costa Rica, Ecuador, Honduras, and Nicaragua.

Code information

Reagent lot numbers: A7125, exp. date 01/9/2019; A7126, exp. date 01/11/2019; A7127, exp. date 01/09/2019, A7128, exp. 01/15/2019; A7129, exp. 01/15/2019; A7130, exp. 01/17/2019; A7131, exp. 01/17/2019, A8003, exp. 01/31/2019; A8004, exp. 01/31/2019; A8005, exp. 02/04/2019; and A8006, exp. 02/04/2019.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2953-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Sysmex America, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2953-2018. Report a data issue: hello@regdatalab.com.