Z-1717-2020 Class II Terminated

Recalled by Sysmex America, Inc. — Mundelein, IL

Recall Details

Product Type
Devices
Report Date
April 22, 2020
Initiation Date
January 21, 2020
Termination Date
January 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product Description

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reason for Recall

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Distribution Pattern

USA: FL, VT, IL.

Code Information

Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620