Z-1717-2020 Class II Terminated

Recalled by Sysmex America, Inc. — Mundelein, IL

FDA device recall Z-1717-2020 was initiated by Sysmex America, Inc. on Jan 21, 2020 and is designated Class II. Reason for recall: When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a… The recall status is terminated (terminated Jan 30, 2023). Affected quantity: 40 units.

Recall details

Product Type
Devices
Report Date
Apr 22, 2020
Initiation Date
Jan 21, 2020
Termination Date
Jan 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product description

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reason for recall

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Distribution pattern

USA: FL, VT, IL.

Code information

Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1717-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1717-2020. Report a data issue: hello@regdatalab.com.