Z-1722-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 28, 2023
- Initiation Date
- May 10, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Product Description
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Reason for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Distribution Pattern
US Nationwide - Worldwide Distribution
Code Information
GTIN 00763000615611. Lot Serial Numbers: RSL602260S, RSL602263S, RSL602265S