Z-1729-2019 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-1729-2019 was initiated by Leica Microsystems, Inc. on May 3, 2019 and is designated Class II. Reason for recall: The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious injuries. The recall status is terminated (terminated Sep 28, 2023). Affected quantity: 17,179 devices total (4,835 US, 12,344 OUS).

Recall details

Product Type
Devices
Report Date
Jun 12, 2019
Initiation Date
May 3, 2019
Termination Date
Sep 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17,179 devices total (4,835 US, 12,344 OUS)

Product description

Leica Cryostats, used to produce frozen sample sections for biological, medical and industrial applications, including in-vitro diagnostic (IVD) applications.

Reason for recall

The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious injuries.

Distribution pattern

Distributed to Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District Of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming. Distributed internationally to Canada.

Code information

Leica CM1950 - All devices with serial number: 8092; 8091; 8089 to 8087; 8081; 8078; 8077; 8073; 8071 to 8067; 8064 to 8062; 8058 to 8053; 8051 to 8049; 8047; 8043; 8040; 8038 to 8033; 8031 to 8025; 8023 to 8004; 8002 to 7940; 7938 to 7892; 7890 to 7469; 7466; and all devices with serial number lower than 7466. Leica CM1860/UV - All devices with serial number: 4201; 4200;4196; 4193; 4192; 4188; 4186 to 4184; 4180; 4179; 4173 to 4170; 4168; 4166 to 4160; 4158; 4157; 4154 to 4152; 4149; 4147; 4145 to 4135; 4133 to 4092; 4090; and all devices with serial number lower than 4090. Leica CM1520 - All devices with serial number: 1450; 1448; 1446 to 1444; 1442; 1439; 1437; and all devices with serial number lower than 1437. Leica CM3050 S - All devices with serial number: 7398; 7396 to 7394; 7392; 7398 to 7387; 7385 to 7381; 7379; and all devices with serial number lower than 7379. Leica CM1850UV - 2822; and all devices with serial number lower than 2822. Leica CM1850 - 7265; and all devices with serial number lower than 7265. Leica CM1510 S - 6600; and all devices with serial number lower than 6600. Leica CM1100 - 1418; and all devices with serial number lower than 1418

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1729-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Leica Microsystems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Leica Microsystems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Leica Microsystems, Inc. →

Source: openFDA record for Z-1729-2019. Report a data issue: hello@regdatalab.com.