Z-1734-2023 Class I Ongoing

Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) — Mounds View, MN

Recall Details

Product Type
Devices
Report Date
June 28, 2023
Initiation Date
May 10, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product Description

CRTD VIVA QUAD XT CUST PARY GOLD, Model Number DTBA1QQPG; Implantable Cardioverter Defibrillators

Reason for Recall

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Distribution Pattern

US Nationwide - Worldwide Distribution

Code Information

GTIN 00643169968585, Lot Serial Numbers: BLC229283H, BLC229226H, BLC229270H, BLC229280H, BLC229229H, BLC228997H, BLC229238H, BLC229043H, BLC229231H, BLC229046H, BLC229045H, BLC229230H, BLC228995H, BLC229010H