Z-1749-2015 Class II Terminated
FDA device recall Z-1749-2015 was initiated by Orthofix, Inc on May 18, 2015 and is designated Class II. Reason for recall: The Firebird Spinal Fixation System 6.5 mm x 55 mm Bone Screw, Self-Tapping is currently mislabeled at the distal tip of the bone screw showing a length measurement of 50 mm instead of 55 mm. The recall status is terminated (terminated Oct 30, 2015). Affected quantity: 45 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 17, 2015
- Initiation Date
- May 18, 2015
- Termination Date
- Oct 30, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 45 units
Product description
Reason for recall
Distribution pattern
US Distribution to the states of : OK, MD, CA, FL, MI, SC, CO, NY and PA., and Internationally to Asturias Spain.