Z-1761-2020 Class II Terminated

Recalled by Radiometer Medical ApS — Bronshoj, N/A

FDA device recall Z-1761-2020 was initiated by Radiometer Medical ApS on Feb 3, 2020 and is designated Class II. Reason for recall: The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in… The recall status is terminated (terminated Sep 20, 2022). Affected quantity: 3,968 units.

Recall details

Product Type
Devices
Report Date
Apr 29, 2020
Initiation Date
Feb 3, 2020
Termination Date
Sep 20, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,968 units

Product description

ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for recall

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

Distribution pattern

Worldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada, Netherlands, China, Croatia, Czech Republic, Denmark, Germany, Spain, France, Hungary, India, Japan, South Korea, United Kingdom, Poland, Australia, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland.

Code information

All software versions below 3.4 MR2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1761-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Radiometer Medical ApS? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Radiometer Medical ApS: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Radiometer Medical ApS →

Source: openFDA record for Z-1761-2020. Report a data issue: hello@regdatalab.com.