Z-1765-2017 Class II Terminated

Recalled by ORTHO-CLINICAL DIAGNOSTICS — Bridgend, N/A

FDA device recall Z-1765-2017 was initiated by ORTHO-CLINICAL DIAGNOSTICS on Mar 7, 2017 and is designated Class II. Reason for recall: Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have… The recall status is terminated (terminated Jun 14, 2019). Affected quantity: US: 6324; Foreign: 1868.

Recall details

Product Type
Devices
Report Date
Apr 19, 2017
Initiation Date
Mar 7, 2017
Termination Date
Jun 14, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 6324; Foreign: 1868

Product description

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Reason for recall

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

Distribution pattern

Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.

Code information

Lot 1570 (expiry 22 March 2017), Lot 1580 (expiry 13 April 2017), Lot 1590 (expiry 27 April 2017)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1765-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ORTHO-CLINICAL DIAGNOSTICS? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for ORTHO-CLINICAL DIAGNOSTICS: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about ORTHO-CLINICAL DIAGNOSTICS →

Source: openFDA record for Z-1765-2017. Report a data issue: hello@regdatalab.com.