Z-1772-2022 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-1772-2022 was initiated by Randox Laboratories Ltd. on Aug 1, 2022 and is designated Class II. Reason for recall: ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg… The recall status is terminated (terminated Dec 13, 2024). Affected quantity: 2 units.

Recall Details

Product Type
Devices
Report Date
Sep 28, 2022
Initiation Date
Aug 1, 2022
Termination Date
Dec 13, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Reason for Recall

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Distribution Pattern

US Nationwide distribution in the states of WV, NC.

Code Information

GTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1772-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.