Z-1775-2019 Class II Terminated
FDA device recall Z-1775-2019 was initiated by Conformis, Inc. on Mar 29, 2019 and is designated Class II. Reason for recall: Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants. The recall status is terminated (terminated Apr 28, 2020). Affected quantity: 128.
Recall details
- Product Type
- Devices
- Report Date
- Jun 19, 2019
- Initiation Date
- Mar 29, 2019
- Termination Date
- Apr 28, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 128
Product description
Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)
Reason for recall
Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.
Distribution pattern
Nationwide
Code information
HBS-033-0010-010101: Serial Numbers:423398, 424244, 424392, 425455, 425607, 425755, 426309, 427186, 428901, 430404,430766, 431394,431844, 434871, 436520, 438363, 438925, 439064, 439561, 440369, 441662, 441734; HBS-033-0011-010101: Serial Numbers:422796, 424450, 424510, 426034, 426118, 430150, 430534, 431142, 431502, 431581, 432030, 433555, 434136, 434439, 435282, 436108, 436212, 437452, 437657, 437745, 437968, 440603, 442816; HBS-033-0012-010101: Serial Numbers: 425346, 425608, 425866, 427702, 427754, 429459, 429844, 430053, 430087, 432011, 432363, 436792, 437653, 438516, 439063, 440253; HBS-033-0013-010101: Serial Numbers: 422797, 424647, 425456, 426297, 427360, 427898, 428372, 428957, 429014, 429921, 429937, 430338, 430613, 430962, 433445, 434149, 434451, 434635, 435118, 436125, 436230, 436333, 437073, 438024, 438112, 439641, 439929, 441028, 441030; HBS-033-0014-010101: Serial Numbers: 423231, 425276, 426069, 426382, 427480, 427752, 427883, 428820, 428900, 431315, 431333, 431977, 432543, 432953, 434612, 435649, 435883, 435910, 436122, 437322, 437863, 438631, 440445; HBS-033-0015-010101: Serial Numbers: 429552, 432936, 433085, 437373, 437681, 437925, 439272 HBS-033-0016-010101 425418, 428545, 431199, 435657, 436790, 437425, 437656, 442745; HBS-033-0016-010101: Serial Numbers:425418, 428545, 431199, 435657, 436790, 437425, 437656, 442745
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1775-2019
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026