Z-1776-2019 Class II Terminated

Recalled by Conformis, Inc. — Billerica, MA

FDA device recall Z-1776-2019 was initiated by Conformis, Inc. on Mar 29, 2019 and is designated Class II. Reason for recall: Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants. The recall status is terminated (terminated Apr 28, 2020). Affected quantity: 183.

Recall details

Product Type
Devices
Report Date
Jun 19, 2019
Initiation Date
Mar 29, 2019
Termination Date
Apr 28, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
183

Product description

Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)

Reason for recall

Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.

Distribution pattern

Nationwide

Code information

HBS-033-0010-020101: Serial Numbers: 425365, 425428, 425767, 425954, 426249, 426408, 426508, 426722, 426725, 428146, 428328, 429186, 429310, 429857, 430590, 431869, 432271, 432292, 432555, 432773, 433524, 434063, 436063, 436522, 438792, 439464, 439559, 441029; HBS-033-0011-020101: Serial Numbers: 423937, 424879, 425273, 425425, 425457, 425605, 425627, 426730, 427699, 427760, 427897, 428066, 428791, 428802, 429699, 429829, 429856, 429940, 430691, 431029, 432687, 434510, 435069, 435409, 435886, 435961, 436017, 436659, 437864, 437866, 438568, 439309, 439457, 439650, 441929; HBS-033-0012-020101: Serial Numbers:424288, 424377, 426081, 426298, 426581, 426608, 426873, 427353, 427759, 428504, 428744, 428814, 428940, 429787, 429854, 429928, 430045, 430344, 430780, 431025, 431032, 431443, 431813, 432098, 432925, 434265, 434704, 435120, 435483, 435790, 436682, 436891, 437036, 437050, 438471, 438511, 438570, 439453, 439471, 439738, 441009, 441527, 442347; HBS-033-0013-020101: Serial Numbers: 423539, 425277, 426121, 430033, 430129, 430461, 432012, 434291, 435938, 436951, 437614, 437654, 437776, 437932, 438157, 438304, 438384, 439822, 440273, 440446, 441053, 441714, 442149; HBS-033-0014-020101: Serial Numbers: 423852, 425345, 425968, 426897, 427830, 428294, 428697, 428836, 429416, 430028, 431541, 432156, 432399, 432421, 432440, 432524, 433245, 434239, 434884, 435146, 435366, 437208, 438079, 438485, 438916, 438991, 439849, 440604, 440673, 441492; HBS-033-0015-020101: Serial Numbers: 426179, 429192, 429828, 429855, 431489, 431512, 432017, 433642, 436330, 437089, 437495, 437651, 437714, 437906, 438105, 438296, 438739, 438784, 440265; HBS-033-0016-020101: Serial Numbers: 426119, 435562, 437516, 438708, 440704

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1776-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Conformis, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1776-2019. Report a data issue: hello@regdatalab.com.