Z-1783-2014 Class II Terminated
FDA device recall Z-1783-2014 was initiated by Philips Healthcare Inc. on Mar 5, 2013 and is designated Class II. Reason for recall: If a customer creates customized trend scales in the trend review tile and the iX or primary server reboots for any reason, the iX(s) will enter a reboot loop. If one iX reboots, the issue will… The recall status is terminated (terminated May 17, 2017). Affected quantity: 899.
Recall details
- Product Type
- Devices
- Report Date
- Jun 18, 2014
- Initiation Date
- Mar 5, 2013
- Termination Date
- May 17, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 899
Product description
Reason for recall
Distribution pattern
Worldwide Distribution: US (nationwide) and country of: Canada.