Z-1791-2018 Class II Terminated

Recalled by TP Orthodontics, Inc. — La Porte, IN

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
May 1, 2018
Termination Date
April 15, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Jasper Vektor, Class II Correction Appliance, Part No. 610-533Lused for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

Reason for Recall

The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.

Distribution Pattern

Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.

Code Information

17515000, 25915000