Z-1793-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- May 1, 2018
- Termination Date
- April 15, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5
Product Description
Jasper Vektor, Class II Correction Appliance, Part No. 610-533R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.
Reason for Recall
The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.
Distribution Pattern
Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.
Code Information
17515000, 17615000, 25915000