Z-1793-2020 Class II Terminated
Recalled by bioMerieux, Inc.
— Durham, NC
FDA device recall Z-1793-2020 was initiated by bioMerieux, Inc. on Jan 24, 2020 and is designated Class II. Reason for recall: The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument. The recall status is terminated (terminated Jan 11, 2023).
Recall details
- Product Type
- Devices
- Report Date
- Apr 29, 2020
- Initiation Date
- Jan 24, 2020
- Termination Date
- Jan 11, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 Plus printers) PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).
Reason for recall
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.
Distribution pattern
Worldwide distribution. US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.
Code information
Instrument Serial Number AS180-00028 1035196 AS180-00067 1007580 AS180-00084 2055833 AS180-00105 2025461 AS180-00138 2025461 AS180-00179 1039938 AS180-00217 1033454 AS180-00248 1029262 AS180-00278 1031886 AS180-00300 1033400 AS180-00306 1034034 AS180-00326 1034190 AS180-00336 1041241 AS180-00339 2054793 AS180-00357 2055775 AS180-00381 1073849 AS180-00385 1054234 AS180-00398 1052053 AS180-00402 2055620 AS180-00405 1063739 AS180-00418 1042188 AS180-00420 1040343 AS180-00428 1030890 AS180-00453 1031472 AS180-00468 1043640 AS180-00476 1029272 AS180-00478 1053432 AS180-00479 1011691 AS180-00480 5001554 AS180-00481 1039588 AS180-00482 1034363 AS180-00483 1029990 AS180-00484 1049298 AS180-00485 AU01 AS180-00486 1047965 AS180-00487 1074074 AS180-00488 1000283 AS180-00489 1035594 AS180-00490 1047965 AS180-00491 1035549 AS180-00492 1070878 AS180-00493 1090075 AS180-00494 1039290 AS180-00495 1029148 AS180-00500 1007884 AS180-00509 1065153
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1793-2020
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for bioMerieux, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for bioMerieux, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about bioMerieux, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works