Z-1795-2023 Class I Ongoing

Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) — Mounds View, MN

Recall Details

Product Type
Devices
Report Date
June 28, 2023
Initiation Date
May 10, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product Description

ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators

Reason for Recall

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Distribution Pattern

US Nationwide - Worldwide Distribution

Code Information

GTIN 00643169720466, Lot Serial Numbers: BWI218211H, BWI218218H, BWI218226H, BWI218228H, BWI218206H, BWI218224H, BWI218210H, BWI218222H, BWI218217H, BWI218214H, BWI218216H, BWI218208H, BWI218207H, BWI218221H, BWI218212H, BWI218220H