Z-1815-2022 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
October 5, 2022
Initiation Date
August 12, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
44 units (5/box); 133 units (5/box) OUS

Product Description

Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.

Reason for Recall

An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

Distribution Pattern

Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA

Code Information

UDI-DI: 00821925044579 Lot Numbers: CA191182 CA191183 CA191184 CA191185 CA191186 CA236235 CA236237 FR230559