Z-1819-2015 Class II Terminated

Recalled by Orthofix, Inc — Lewisville, TX

FDA device recall Z-1819-2015 was initiated by Orthofix, Inc on May 11, 2015 and is designated Class II. Reason for recall: There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting. The recall status is terminated (terminated Feb 17, 2016). Affected quantity: 36 units.

Recall details

Product Type
Devices
Report Date
Jul 1, 2015
Initiation Date
May 11, 2015
Termination Date
Feb 17, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36 units

Product description

ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.

Reason for recall

There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.

Distribution pattern

Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.

Code information

All lot codes

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1819-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Orthofix, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1819-2015. Report a data issue: hello@regdatalab.com.