Z-1821-2012 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-1821-2012 was initiated by Philips Healthcare Inc. on May 11, 2012 and is designated Class II. Reason for recall: If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally… The recall status is terminated (terminated Oct 6, 2016). Affected quantity: 6.

Recall details

Product Type
Devices
Report Date
Jun 27, 2012
Initiation Date
May 11, 2012
Termination Date
Oct 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6

Product description

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Reason for recall

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

Distribution pattern

6 US customers

Code information

All systems with Eleva software version 4.x and stitching option.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1821-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-1821-2012. Report a data issue: hello@regdatalab.com.