FDA device recall Z-1825-2020 was initiated by Elekta Inc on Apr 16, 2020 and is designated Class II. Reason for recall: The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose. The recall status is ongoing. Affected quantity: 96 devices.
Recall details
- Product Type
- Devices
- Report Date
- May 6, 2020
- Initiation Date
- Apr 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 96 devices
Product description
Elekta MONACO RTP Sytem, radiation treatment planning software system - Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The
Reason for recall
The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.
Distribution pattern
Worldwide distribution - US Nationwide Distribution in the states of IA, NJ, NY, PA, TX, WI and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, United Kingdom.
Code information
Software Build: 5.40.00 (UDI: (01)00858164002190(10) 5.40.00), 5.40.01 (UDI: 01)00858164002190(10) 5.40.01)