Z-1825-2026 Class II Ongoing

Recalled by PHILIPS MEDICAL SYSTEMS — Cambridge, MA

FDA device recall Z-1825-2026 was initiated by PHILIPS MEDICAL SYSTEMS on Mar 7, 2026 and is designated Class II. Reason for recall: Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential… The recall status is ongoing. Affected quantity: 3 systems.

Recall details

Product Type
Devices
Report Date
Apr 22, 2026
Initiation Date
Mar 7, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 systems

Product description

Philips Spectral CT on Rails. Model Number: 728334.

Reason for recall

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Distribution pattern

Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.

Code information

Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1825-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for PHILIPS MEDICAL SYSTEMS? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1825-2026. Report a data issue: hello@regdatalab.com.