Z-2652-2026 Class II Ongoing

Recalled by PHILIPS MEDICAL SYSTEMS — Cambridge, MA

FDA device recall Z-2652-2026 was initiated by PHILIPS MEDICAL SYSTEMS on June 11, 2026 and is designated Class II. Reason for recall: The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acqui… The recall status is ongoing. Affected quantity: 182.

Recall Details

Product Type
Devices
Report Date
August 5, 2026
Initiation Date
June 11, 2026
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
182

Product Description

Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil

Reason for Recall

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

Distribution Pattern

US Nationwide Distribution including Puerto Rico

Code Information

FCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325