Z-2652-2026 Class II Ongoing
FDA device recall Z-2652-2026 was initiated by PHILIPS MEDICAL SYSTEMS on Jun 11, 2026 and is designated Class II. Reason for recall: The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical… The recall status is ongoing. Affected quantity: 182.
Recall details
- Product Type
- Devices
- Report Date
- Aug 5, 2026
- Initiation Date
- Jun 11, 2026
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 182
Product description
Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil
Reason for recall
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
Distribution pattern
US Nationwide Distribution including Puerto Rico
Code information
FCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325