FDA device recall Z-1826-2020 was initiated by Elekta Inc on Mar 30, 2020 and is designated Class II. Reason for recall: A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System. The recall status is terminated (terminated Apr 10, 2024). Affected quantity: 25 units.
Recall details
- Product Type
- Devices
- Report Date
- May 6, 2020
- Initiation Date
- Mar 30, 2020
- Termination Date
- Apr 10, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25 units.
Product description
Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument
Reason for recall
A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.
Distribution pattern
US nationwide distributions in the states of MA, NH, NC, UT, OH, FL, MN, NY, LA and MT.
Code information
Serial Numbers:SH00106, SH00120, SH00149, SH00160, SH00165, SH00110, SH00143, SH00150, SH00161, SH00166, SH00111, SH00144, SH00157, SH00162, SH00167, SH00112, SH00147, SH00158, SH00163, SH00168, SH00114, SH00148, SH00159, SH00164 and SH00178. UDI 7340048304887.