Z-1828-2017 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-1828-2017 was initiated by Merge Healthcare, Inc. on Apr 4, 2016 and is designated Class II. Reason for recall: During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently. The recall status is terminated (terminated Jun 22, 2020). Affected quantity: 1,627 sites potentially have the affected versions.

Recall details

Product Type
Devices
Report Date
Apr 26, 2017
Initiation Date
Apr 4, 2016
Termination Date
Jun 22, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,627 sites potentially have the affected versions

Product description

Merge Eye Station and Merge Eye Care PACS.

Reason for recall

During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently.

Distribution pattern

Distribution was nationwide to medical facilities. Foreign distribution was made to Canada, as well as to other countries. There was also government distribution.

Code information

All versions