Z-1829-2015 Class II Terminated

Recalled by Varian Medical Systems Inc — Charlottesville, VA

FDA device recall Z-1829-2015 was initiated by Varian Medical Systems Inc on May 18, 2015 and is designated Class II. Reason for recall: Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06… The recall status is terminated (terminated Jan 26, 2016). Affected quantity: 46.

Recall details

Product Type
Devices
Report Date
Jul 1, 2015
Initiation Date
May 18, 2015
Termination Date
Jan 26, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
46

Product description

Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.

Reason for recall

Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat

Distribution pattern

Worldwide Distribution.

Code information

JAQ

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1829-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Varian Medical Systems Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1829-2015. Report a data issue: hello@regdatalab.com.