Z-1836-2020 Class II Ongoing

Recalled by Radiometer Medical ApS — Bronshoj, N/A

Recall Details

Product Type
Devices
Report Date
May 6, 2020
Initiation Date
March 6, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

ABL800 FLEX model number 393-800 and 393-801, UDI:(01)05700693938004

Reason for Recall

Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.

Distribution Pattern

US: in the state of California

Code Information

Serial Numbers: 902-754R2692N002 and 902-754R2692N001