Z-1865-2019 Class II Terminated

Recalled by Thermo Fisher Scientific (Asheville) LLC — Marietta, OH

FDA device recall Z-1865-2019 was initiated by Thermo Fisher Scientific (Asheville) LLC on April 2, 2019 and is designated Class II. Reason for recall: Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by th… The recall status is terminated (terminated July 23, 2020). Affected quantity: 10 units.

Recall Details

Product Type
Devices
Report Date
July 3, 2019
Initiation Date
April 2, 2019
Termination Date
July 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 units

Product Description

Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05

Reason for Recall

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Distribution Pattern

CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia

Code Information

Serial Numbers: 30024679S 300246796 300246797 300246798 300246799 300247393 300247394 30024739S 300247396 3002477S8