Z-1870-2019 Class II Terminated

Recalled by Thermo Fisher Scientific (Asheville) LLC — Marietta, OH

FDA device recall Z-1870-2019 was initiated by Thermo Fisher Scientific (Asheville) LLC on April 2, 2019 and is designated Class II. Reason for recall: Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by th… The recall status is terminated (terminated July 23, 2020). Affected quantity: 4 units.

Recall Details

Product Type
Devices
Report Date
July 3, 2019
Initiation Date
April 2, 2019
Termination Date
July 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

Thermo Scientific General Purpose Dynamic Bath Precision 2L-Shallow GP Bath Model Number: TSGP2S

Reason for Recall

Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

Distribution Pattern

CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia

Code Information

Serial Numbers: 300246821 300246822 300246823 300247426