Z-1887-2017 Class II Terminated

Recalled by Ortho-Clinical Diagnostics — Rochester, NY

FDA device recall Z-1887-2017 was initiated by Ortho-Clinical Diagnostics on Feb 21, 2017 and is designated Class II. Reason for recall: Potential for biased results generated using the specific lots of VITROS VANC Reagent. This can occur after the reagent packs are loaded onto the analyzer and stored within the on-analyzer stability… The recall status is terminated (terminated Jun 19, 2019). Affected quantity: US: 2740; OUS: 616.

Recall details

Product Type
Devices
Report Date
May 3, 2017
Initiation Date
Feb 21, 2017
Termination Date
Jun 19, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 2740; OUS: 616

Product description

VITROS Chemistry Products VANC Reagent, REF/Product Code 680 1709, Unique Identifier No. 10758750006731, Rx ONLY, IVD --- Ortho Clinical Diagnostics - USA -- Ortho-Clinical Diagnostics Pencoed, CF35 5PZ, UK ---- The VITROS VANC reagent is for ORTHO Clinical Diagnostics manufactured by Siemens. The VANC product has a maximum on-analyzer stability of < 7days. The on-analyzer stability of each VITROS VANC Reagent pack is determined by the system. Once the VANC reagent pack is loaded on the system, the on-analyzer stability time is continuously adjusted based upon both the amount of time the reagent pack is on the analyzer and the number of tests remaining in the pack. The method by which the on-analyzer stability of the VITROS VANC Reagent is determined is stated in the reagents Instructions for Use.

Reason for recall

Potential for biased results generated using the specific lots of VITROS VANC Reagent. This can occur after the reagent packs are loaded onto the analyzer and stored within the on-analyzer stability date.

Distribution pattern

Worldwide Distribution - US Distribution including Puerto Rico and to the countries of : Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Spain Singapore and Venezuela.

Code information

Lot numbers: 31-5407 (exp. date 2017-05-23), 31-5474 (exp. date 2017-07-04), 31-5506 (exp. date 2017-07-04)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1887-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ortho-Clinical Diagnostics? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ortho-Clinical Diagnostics: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ortho-Clinical Diagnostics →

Source: openFDA record for Z-1887-2017. Report a data issue: hello@regdatalab.com.