Z-1920-2025 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1920-2025 was initiated by Olympus Corporation Of The Americas on May 6, 2025 and is designated Class II. Reason for recall: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to… The recall status is ongoing. Affected quantity: 201 units (100 US, 101 OUS).

Recall details

Product Type
Devices
Report Date
Jun 18, 2025
Initiation Date
May 6, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
201 units (100 US, 101 OUS)

Product description

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number: TFL-SLS, TFL-SLS-JA (Japan only); Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No

Reason for recall

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Code information

All lot/serial numbers for the following products: TFL-SLS UDI: 00821925044135; TFL-SLS-JA (Japan Only)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1920-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1920-2025. Report a data issue: hello@regdatalab.com.