Z-1923-2017 Class II Terminated
FDA device recall Z-1923-2017 was initiated by Merge Healthcare, Inc. on Mar 30, 2015 and is designated Class II. Reason for recall: The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias… The recall status is terminated (terminated May 5, 2017). Affected quantity: 23 sensors (PhaseIn Units).
Recall details
- Product Type
- Devices
- Report Date
- May 10, 2017
- Initiation Date
- Mar 30, 2015
- Termination Date
- May 5, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 23 sensors (PhaseIn Units)
Product description
Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
Reason for recall
The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
Distribution pattern
US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
Code information
ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841.