Z-1929-2020 Class II Ongoing

Recalled by Encore Medical, LP — Austin, TX

Recall Details

Product Type
Devices
Report Date
May 20, 2020
Initiation Date
October 22, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
86 units

Product Description

EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.

Reason for Recall

The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.

Distribution Pattern

US Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.

Code Information

All lots