Z-1950-2012 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 18, 2012
- Initiation Date
- July 2, 2012
- Termination Date
- July 1, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 65
Product Description
Drager PT 4000 Phototherapy System
Reason for Recall
A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b
Distribution Pattern
Nationwide Distribution-USA (nationwide) and Puerto Rico.
Code Information
Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.