Z-1978-2019 Class II Terminated

Recalled by CooperSurgical, Inc. — Trumbull, CT

FDA device recall Z-1978-2019 was initiated by CooperSurgical, Inc. on Jun 14, 2019 and is designated Class II. Reason for recall: The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as… The recall status is terminated (terminated May 19, 2020). Affected quantity: 213 units in total.

Recall details

Product Type
Devices
Report Date
Jul 24, 2019
Initiation Date
Jun 14, 2019
Termination Date
May 19, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
213 units in total

Product description

Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30 The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).

Reason for recall

The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.

Distribution pattern

US distribution to state of: AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..

Code information

Lot 021-19

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1978-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for CooperSurgical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for CooperSurgical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about CooperSurgical, Inc. →

Source: openFDA record for Z-1978-2019. Report a data issue: hello@regdatalab.com.