Z-1980-2019 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

Recall Details

Product Type
Devices
Report Date
July 24, 2019
Initiation Date
June 24, 2019
Termination Date
April 1, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Reason for Recall

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

Distribution Pattern

US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Code Information

UDI/GTIN: 5060081071321 Installed product: UNITY SYSTEM/13553-006/600003 UNITY SYSTEM/10817-011/600010 UNITY SYSTEM/12473-020/600020 UNITY SYSTEM/12274-017/600012 UNITY SYSTEM/10562-011/600016 UNITY SYSTEM/10719-T03001/600013 UNITY SYSTEM/30004070-001/600023 UNITY SYSTEM/10420-AVL-U/600008 UNITY SYSTEM/11014-UMCU-U/600007 UNITY SYSTEM/10157-006/600014 UNITY SYSTEM/11611-40/600011 UNITY SYSTEM/11489-21/600009