Z-1990-2023 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1990-2023 was initiated by Olympus Corporation Of The Americas on May 4, 2023 and is designated Class II. Reason for recall: Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms… The recall status is ongoing. Affected quantity: 3795 pieces.

Recall details

Product Type
Devices
Report Date
Jun 28, 2023
Initiation Date
May 4, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3795 pieces

Product description

Single Use Repositionable Clip QUICKCLIP PRO 230CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.B

Reason for recall

Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI: 04953170385940 Lot Numbers: 24K, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1990-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Olympus Corporation Of The Americas? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Olympus Corporation Of The Americas: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Olympus Corporation Of The Americas →

Source: openFDA record for Z-1990-2023. Report a data issue: hello@regdatalab.com.