Z-1995-2020 Class I Terminated
Recalled by Medtronic Navigation, Inc.
— Louisville, CO
FDA device recall Z-1995-2020 was initiated by Medtronic Navigation, Inc. on Aug 9, 2019 and is designated Class I. Reason for recall: Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and… The recall status is terminated (terminated Apr 2, 2024). Affected quantity: 61 customers purchased the specified combination of equipment which may lead to this error.
Recall details
- Product Type
- Devices
- Report Date
- May 27, 2020
- Initiation Date
- Aug 9, 2019
- Termination Date
- Apr 2, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 61 customers purchased the specified combination of equipment which may lead to this error.
Product description
Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.
Reason for recall
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
Distribution pattern
Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom.
Code information
Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS procedure. Stealthware Station Cranial Software GTIN(s): 00643169878204, 00643169625549 O-Arm Imaging System GTIN(s): 00763000068073, 00763000068080, 00763000068097, 00763000068103, 00763000074050, 00763000074067, 00763000074074, 00763000074081, 00763000074098, 00763000074104, 00763000074111, 00763000074128, 00763000074135, 00763000079000, 00763000079017, 00763000079024, 00763000079031, 00763000079048, 00763000079055, 00763000079062, 00763000079079, 00763000079086, 00763000079093, 00763000079109, 00763000079215, 00763000079246, 00763000079260, 00763000080969, 00763000080976, 00763000080983, 00763000080990, 00763000081003, 00763000081010, 00763000081027, 00763000081034, 00763000081041, 00763000081058, 00763000081065, 00763000081072, 00763000081089, 00763000081096 and 00763000081102. Nexframe Stereotactic System GTIN(s): DB-2040, 00643169214811; MI-1000, 00643169986763; MI-2000, 00643169986770; NP-1000, 00613994747877.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1995-2020
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Medtronic Navigation, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works