Z-1995-2020 Class I Terminated

Recalled by Medtronic Navigation, Inc. — Louisville, CO

FDA device recall Z-1995-2020 was initiated by Medtronic Navigation, Inc. on Aug 9, 2019 and is designated Class I. Reason for recall: Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and… The recall status is terminated (terminated Apr 2, 2024). Affected quantity: 61 customers purchased the specified combination of equipment which may lead to this error.

Recall details

Product Type
Devices
Report Date
May 27, 2020
Initiation Date
Aug 9, 2019
Termination Date
Apr 2, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
61 customers purchased the specified combination of equipment which may lead to this error.

Product description

Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.

Reason for recall

Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).

Distribution pattern

Worldwide distribution including US nationwide, Austria, Canada, Italy, Romania, Spain, and United Kingdom.

Code information

Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS procedure. Stealthware Station Cranial Software GTIN(s): 00643169878204, 00643169625549 O-Arm Imaging System GTIN(s): 00763000068073, 00763000068080, 00763000068097, 00763000068103, 00763000074050, 00763000074067, 00763000074074, 00763000074081, 00763000074098, 00763000074104, 00763000074111, 00763000074128, 00763000074135, 00763000079000, 00763000079017, 00763000079024, 00763000079031, 00763000079048, 00763000079055, 00763000079062, 00763000079079, 00763000079086, 00763000079093, 00763000079109, 00763000079215, 00763000079246, 00763000079260, 00763000080969, 00763000080976, 00763000080983, 00763000080990, 00763000081003, 00763000081010, 00763000081027, 00763000081034, 00763000081041, 00763000081058, 00763000081065, 00763000081072, 00763000081089, 00763000081096 and 00763000081102. Nexframe Stereotactic System GTIN(s): DB-2040, 00643169214811; MI-1000, 00643169986763; MI-2000, 00643169986770; NP-1000, 00613994747877.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1995-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-1995-2020. Report a data issue: hello@regdatalab.com.