Z-1996-2016 Class II Terminated
FDA device recall Z-1996-2016 was initiated by Ethicon, Inc. on Apr 11, 2016 and is designated Class II. Reason for recall: Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid… The recall status is terminated (terminated Aug 2, 2017). Affected quantity: 228 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 22, 2016
- Initiation Date
- Apr 11, 2016
- Termination Date
- Aug 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 228 units
Product description
EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with manual compression as an adjunct to hemostasis for control of bleeding during adult liver surgery and soft tissue bleeding in adults when control of bleeding by standard surgical methods of hemostasis is ineffective or impractical.
Reason for recall
Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date (2016-12-28).
Distribution pattern
Nationwide Distribution
Code information
Product Code #EVT5024, Lot #U01F012A