Z-1997-2026 Class III Ongoing

Recalled by Heraeus Medical GmbH (Dental Division) — Wehrheim

FDA device recall Z-1997-2026 was initiated by Heraeus Medical GmbH (Dental Division) on Mar 25, 2026 and is designated Class III. Reason for recall: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process… The recall status is ongoing. Affected quantity: 51,771 units.

Recall details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
Mar 25, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
51,771 units

Product description

Heraeus, PALAMIX uno. Material Number: 66057893.

Reason for recall

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;

Code information

Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1997-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1997-2026. Report a data issue: hello@regdatalab.com.