Z-1998-2026 Class III Ongoing

Recalled by Heraeus Medical GmbH (Dental Division) — Wehrheim

Recall Details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
March 25, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,125 units

Product Description

Heraeus, PALAMIX duo. Material Number: 66057897.

Reason for Recall

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;

Code Information

Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;