Z-1998-2026 Class III Ongoing
FDA device recall Z-1998-2026 was initiated by Heraeus Medical GmbH (Dental Division) on Mar 25, 2026 and is designated Class III. Reason for recall: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process… The recall status is ongoing. Affected quantity: 1,125 units.
Recall details
- Product Type
- Devices
- Report Date
- May 6, 2026
- Initiation Date
- Mar 25, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,125 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1998-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab