Z-2022-2020 Class II Terminated
FDA device recall Z-2022-2020 was initiated by K2M, Inc on Apr 13, 2020 and is designated Class II. Reason for recall: The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure. The recall status is terminated (terminated Nov 15, 2021).
Recall details
- Product Type
- Devices
- Report Date
- May 27, 2020
- Initiation Date
- Apr 13, 2020
- Termination Date
- Nov 15, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2022-2020
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026