Z-2049-2017 Class II Terminated
FDA device recall Z-2049-2017 was initiated by Roche Diagnostics Corporation on Mar 21, 2017 and is designated Class II. Reason for recall: Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid… The recall status is terminated (terminated Oct 2, 2017). Affected quantity: 3946.
Recall details
- Product Type
- Devices
- Report Date
- May 17, 2017
- Initiation Date
- Mar 21, 2017
- Termination Date
- Oct 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3946
Product description
Reason for recall
Distribution pattern
US Nationwide.Distribution
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2049-2017
- openFDA file date
- Nov 19, 2026