Z-2050-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 13, 2018
- Initiation Date
- September 17, 2014
- Termination Date
- March 24, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22 licenses
Product Description
nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context.
Reason for Recall
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Distribution Pattern
USA, Sweden, Spain, Russia, Hungary, India, Norway, Iceland, Italy, South Korea, United Kingdom
Code Information
version 2.1 and older