Z-2061-2018 Class II Terminated
FDA device recall Z-2061-2018 was initiated by NordicNeuroLab AS on Jul 3, 2015 and is designated Class II. Reason for recall: The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted… The recall status is terminated (terminated Sep 16, 2020). Affected quantity: 121 licenses.
Recall details
- Product Type
- Devices
- Report Date
- Jun 13, 2018
- Initiation Date
- Jul 3, 2015
- Termination Date
- Sep 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 121 licenses
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia , Belgium, Brazil, China, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway and Poland.