Z-2061-2018 Class II Terminated

Recalled by NordicNeuroLab AS — Bergen, N/A

Recall Details

Product Type
Devices
Report Date
June 13, 2018
Initiation Date
July 3, 2015
Termination Date
September 16, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
121 licenses

Product Description

nordicICE v 2.3.14 nordicICE is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicICE provides both viewing and analysis capabilities of functional and dynamic imaging datasets acquired with MRI or other relevant modalities, including blood oxygen level dependent (BOLD) fMRI, diffusion weighted MRI (DWI) / fiber tracking and dynamic analysis.

Reason for Recall

The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the coregistration.

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia , Belgium, Brazil, China, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway and Poland.

Code Information

Version 2.3.14